SHFA100F SERIES


150W Open-Frame Power Supply for Medical Equipment


  • Wide Operating Voltage 90 to 260 VAC, 47 to 63Hz
  • Single Output
  • Protection: OVP, OLP
  • Size 2"x4"x1.02"
  • Input to Output¡G2MOPP
  • High ESD Immunity
  • Suitable professional healthcare facility

Specifications

 

 

3 Year Warranty

 

Approvals:   

 

Single Output

Model Number

Output Voltage

Output Current

Total Regulation

Max

Output Power

Ripple & Noise

Typ.

Efficiency

Typ. No Load

Consumption

min

max

min (A)

max (A)

(Forced Air)

(mVp-p)

SHFA100-S05F

11.0 VDC

13.0 VDC

11.53

13.63

±3%

150 W

200

86%

0.5 W

SHFA100-S06F

13.0 VDC

16.0 VDC

9.37

11.53

±3%

150 W

300

86%

0.5 W

SHFA100-S07F

16.0 VDC

21.0 VDC

7.14

9.37

±3%

150 W

300

87%

0.5 W

SHFA100-S08F

21.0 VDC

27.0 VDC

5.55

7.14

±3%

150 W

300

88%

0.5 W

SHFA100-S09F

27.0 VDC

33.0 VDC

4.54

5.55

±3%

150 W

300

88%

0.5 W

SHFA100-S10F

33.0 VDC

40.0 VDC

3.75

4.54

±3%

150 W

300

88%

0.5 W

SHFA100-S11F

40.0 VDC

50.0 VDC

3.00

3.75

±3%

150 W

300

88%

0.5 W

 

Electrical Characteristics

Parameter

Test Conditions

Min.

Typ.

Max.

Unit

Safety Approvals Input Voltage Range

Safety Approval & Specification in Label

100

 

240

VAC

Operate Voltage Range

 

90

 

260

VAC

Input Frequency

Sine wave

47

 

63

Hz

Power Factor Correction

 

0.90

 

1

 

Output Power Range

See Rating Chart

 

 

150

W

Low Line Input Current

Full load, Vin=100VAC

 

1.4

 

A

High Line Input Current

Full load, Vin=240VAC

 

0.7

 

A

Low Line Input Inrush Current

Full load, 25°C, Cool start, Vin=100VAC

 

 

50

A

High Line Input Inrush Current

Full load, 25°C, Cool start, Vin=240VAC

 

 

100

A

Safety Ground Leakage Current

Vin=240VAC, Fi=60Hz

 

 

0.25

mA

Efficiency

Full Load, Vin=230VAC

86

 

88

%

Line Regulation

Full Load, Vin=100~120VAC or 200~240VAC

 

 

1

%

Over Voltage Protection

 

112

 

132

%

Over Load Protection

Recovers automatically after fault condition is removed

110

 

150

%

Time of Transient Response

Io=Full Load to Half Load, Vin=110VAC

 

 

4

ms

Hold-Up Time

Full Load, Vin=100VAC

 

 

16

ms

Start Up Time

Full Load, Vin=100~240VAC

 

 

0.5

s

Insulation Resistance

 

50

 

 

MΩ

Ripple & Noise (Peak to Peak)

 

 

 

1

%

Temperature Coefficient

All output

 

 

±0.04

%/°C

Dielectric Withstanding Voltage(P-S)

Primary to Secondary, limit current<10mA

4000

 

 

VAC

Dielectric Withstanding Voltage(P-G)

Primary to PE, limit current<10mA

2828

 

 

VAC

EMC Emission

Compliance to EN55011(CISPR11), EN60601-1-2

B

 

 

Class

 

 

Environmental

Parameter

Test Conditions

Min.

Typ.

Max.

Unit

Operating Temperature

Derate linearly from 100% load at 40℃ to 50% load at 70℃

-10

 

70

°C

Storage Temperature

10~95% RH

-40

 

85

°C

Operating Humidity

non-condensing

0

 

95%

RH

Storage Humidity

 

0

 

95%

RH

Electro Static Discharge

Air Discharge, IEC61000-4-2

 

 

15

KV

Electro Static Discharge

Contact Discharge, IEC61000-4-2

 

 

8

KV

Mean Time Between Failure

Operation Temperature at 25℃J, Calculated per MIL-HDBK-217F

200K

 

 

h

Operating Altitude (Elevation)

All Condition

 

 

3000

m

Vibration

10~500Hz,10min./1cycle, 60min.each along X, Y, Z axes

 

 

5

G

Surge Voltage

  Line-Neutral

 

 

1

KV

Surge Voltage

  Line-PE & Neutral-PE

 

 

2

KV

 

Mechanical Diagram and Technical Chart

 

 

SHFA100F - MEPOS ELECTRONICS
English Danish Dutch French German Italian Japanese Lithuanian Norwegian Portuguese Spanish Swedish

Quotation

No items in cart