Specifications
Â
5 Year Warranty |
 |
Approvals:Â |
Â
Single Output |
||||
Model Number |
Output Voltage |
Max. Output Current |
Total Regulation |
Max. Output Power |
SMDA51-S01 |
3 ~ 5 VDC |
10.0 ~ 8.00 A |
±5% |
40W |
SMDA51-S02 |
5 ~ 6 VDC |
8.00 ~ 6.66 A |
±5% |
40W |
SMDA51-S03 |
6 ~ 8 VDC |
7.00 ~ 5.25 A |
±5% |
42W |
SMDA51-S04 |
 8 ~ 11 VDC |
5.63 ~ 4.09 A |
±4% |
45W |
SMDA51-S05 |
11 ~ 13 VDC |
4.09 ~ 3.46 A |
±3% |
45W |
SMDA51-S06 |
13 ~ 16 VDC |
3.46 ~ 2.81 A |
±3% |
45W |
SMDA51-S07 |
16 ~ 21 VDC |
3.12 ~ 2.38 A |
±3% |
50W |
SMDA51-S08 |
21 ~ 27 VDC |
2.38 ~ 1.85 A |
±2% |
50W |
SMDA51-S09 |
27 ~ 33 VDC |
1.85 ~ 1.51 A |
±2% |
50W |
SMDA51-S10 |
33 ~ 40 VDC |
1.51 ~ 1.25 A |
±2% |
50W |
SMDA51-S11 |
40 ~ 50 VDC |
1.25 ~ 1.00 A |
±2% |
50W |
Â
Â
Multi Output |
|||||||||||||
Part Number |
Output 1 |
Output 2 |
Output 3 |
Max. Output Power |
|||||||||
Vonom |
Iomin |
Iomax |
Regmax |
Vonom |
Iomin |
Iomax |
Regmax |
Vonom |
Iomin |
Iomax |
Regmax |
||
SMDA51-D00 |
+3.3V |
0.5A |
5A |
7% |
+12V |
0.2A |
2A |
5% |
 |
 |
 |
 |
40W |
SMDA51-D01 |
+5V |
0.5A |
5A |
5% |
+12V |
0.2A |
2A |
5% |
 |
 |
 |
 |
42W |
SMDA51-D02 |
+5V |
0.5A |
5A |
5% |
+15V |
0.15A |
1.5A |
5% |
 |
 |
 |
 |
42W |
SMDA51-D03 |
+5V |
0.5A |
5A |
5% |
+24V |
0.1A |
1A |
5% |
 |
 |
 |
 |
42W |
SMDA51-D04 |
+3.3V |
0.5A |
5A |
7% |
+5V |
0.2A |
2A |
5% |
 |
 |
 |
 |
26.5W |
SMDA51-D09 |
+12V |
0.3A |
3A |
5% |
 |
 |
 |
 |
-12V |
0A |
1A |
5% |
42W |
SMDA51-D10 |
+15V |
0.2A |
2A |
5% |
 |
 |
 |
 |
-15V |
0.1A |
1A |
5% |
42W |
SMDA51-D15 |
+5V |
0.5A |
5A |
5% |
 |
 |
 |
 |
-24V |
0.1A |
1A |
5% |
42W |
SMDA51-D16 |
+5.1V |
0.5A |
5A |
5% |
+7.2V |
0.3A |
2.6A |
5% |
 |
 |
 |
 |
42W |
SMDA51-D17 |
+10.5V |
0.2A |
2A |
5% |
 |
 |
 |
 |
-10.5V |
0A |
2A |
5% |
42W |
SMDA51-D18 |
+5V |
0.5A |
5A |
5% |
 |
 |
 |
 |
-5V |
0A |
2A |
5% |
35W |
SMDA51-T00 |
+3.3V |
0.5A |
5A |
7% |
+12V |
0.4A |
2A |
5% |
-12V |
0.1A |
0.8A |
5% |
42W |
SMDA51-T01 |
+5V |
0.5A |
5A |
5% |
+12V |
0.2A |
2A |
5% |
-5V |
0A |
0.8A |
5% |
42W |
SMDA51-T02 |
+5V |
0.5A |
5A |
5% |
+12V |
0.2A |
2A |
5% |
-12V |
0.1A |
0.8A |
5% |
42W |
SMDA51-T03 |
+5V |
0.5A |
5A |
5% |
+15V |
0.4A |
2A |
6% |
-15V |
0.1A |
0.8A |
5% |
42W |
SMDA51-T04 |
+5V |
0.5A |
5A |
5% |
+24V |
0.2A |
1A |
5% |
-24V |
0A |
0.5A |
5% |
42W |
SMDA51-T05 |
+5V |
0.5A |
5A |
5% |
+24V |
0.1A |
1A |
5% |
-12V |
0A |
0.8A |
5% |
42W |
SMDA51-T06 |
+3.3V |
0.5A |
5A |
7% |
+12V |
0.4A |
2A |
5% |
-5V |
0.1A |
0.8A |
5% |
42W |
Â
Â
Â
Electrical Characteristics |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Safety Approvals Input Voltage Range |
Safety Approval & Specification in Label |
100 |
 |
240 |
VAC |
Operate Voltage Range |
Derate linearly from 100% load at 90VAC to 80% load at 80VAC |
80 |
 |
275 |
VAC |
Input Frequency |
Sine wave |
47 |
 |
63 |
Hz |
Output Power Range |
See Rating Chart |
 |
 |
50 |
W |
Low Line Input Current |
Full load, Vin=100VAC |
 |
 |
1.04 |
A |
High Line Input Current |
Full load, Vin=240VAC |
 |
 |
0.53 |
A |
Low Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=100VAC |
 |
 |
35 |
A |
High Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=240VAC |
 |
 |
70 |
A |
Safety Ground Leakage Current |
Vin=264VAC, Fi=63Hz |
 |
 |
0.15 |
mA |
Efficiency |
Full Load, Vin=230VAC |
68 |
 |
83 |
% |
Line Regulation |
Full Load, Vin=100~120VAC or 200~240VAC |
 |
 |
1 |
% |
Over Voltage Protection |
 |
112 |
 |
132 |
% |
Over Load Protection |
Recovers automatically after fault condition is removed |
110 |
 |
150 |
% |
Time of Transient Response |
Io=Full Load to Half Load, Vin=110VAC |
 |
 |
4 |
ms |
Hold-Up Time |
Full Load, Vin=100VAC |
 |
 |
16 |
ms |
Start Up Time |
Full Load, Vin=100~240VAC |
 |
 |
2 |
s |
Insulation Resistance |
 |
50 |
 |
 |
MΩ |
Ripple & Noise (Peak to Peak) |
 |
 |
 |
1 |
% |
Temperature Coefficient |
All output |
 |
 |
±0.04 |
%/°C |
Dielectric Withstanding Voltage(P-S) |
Primary to Secondary, limit current<10mA |
 |
 |
4000 |
VAC |
Dielectric Withstanding Voltage(P-G) |
Primary to PE, limit current<10mA |
 |
 |
1500 |
VAC |
EMC Emission |
Compliance to EN55011(CISPR11), EN60601-1-2 |
B |
 |
 |
Class |
Â
Â
Environmental |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Operating Temperature |
Derate linearly from 100% load at 50â to 50% load at 70â |
-10 |
 |
70 |
°C |
Storage Temperature |
 10~95% RH |
-40 |
 |
85 |
°C |
Operating Humidity |
 non-condensing |
0 |
 |
95% |
RH |
Storage Humidity |
 |
0 |
 |
95% |
RH |
Electro Static Discharge |
Air Discharge, IEC61000-4-2 |
 |
 |
15 |
KV |
Electro Static Discharge |
Contact Discharge, IEC61000-4-2 |
 |
 |
8 |
KV |
Mean Time Between Failure |
Operation Temperature at 25âJ, Calculated per MIL-HDBK-217F |
200K |
 |
 |
h |
Operating Altitude (Elevation) |
All Condition |
 |
 |
3000 |
m |
Vibration |
10~500Hz,10min./1cycle, 60min.each along X, Y, Z axes |
 |
 |
5 |
G |
Surge Voltage |
 Line-Neutral |
 |
 |
1 |
KV |
Surge Voltage |
 Line-PE & Neutral-PE |
 |
 |
2 |
KV |
Â
Â
Derating Chart |
Â
1. Operating Temperature: -10 to 70°C 2. Derate linearly from 100% load at 50°C to 50% load at 70°C   |
Â
Â
Â
Â
Mechanical Drawing |
Â
Â
Â
Datasheet PDF
Â
Â