Specifications
Â
5 Year Warranty |
 |
Approvals:Â |
Â
Single Output |
||||
Product Number |
Output Voltage |
Max. Output Current |
Total Regulation |
Max Output Power |
SMFA120-S02 |
5 ~ 6 Â VDC |
20 ~ 16.66 A |
7% |
100W |
SMFA120-S03 |
6 ~ 9 VDC |
18.33 ~ 12.22 A |
7% |
110W |
SMFA120-S04 |
9 ~ 11 VDC |
13.33 ~ 10.90 A |
5% |
120W |
SMFA120-S05 |
11 ~ 13 VDC |
10.90 ~ 9.23 A |
5% |
120W |
SMFA120-S06 |
13 ~ 16 VDC |
9.23 ~ 7.50 A |
5% |
120W |
SMFA120-S07 |
16 ~ 21 VDC |
7.50 ~ 5.71A |
3% |
120W |
SMFA120-S08 |
21 ~ 27 VDC |
5.71 ~ 4.44 A |
3% |
120W |
SMFA120-S09 |
27 ~ 33 VDC |
4.44 ~ 3.63 A |
2% |
120W |
SMFA120-S10 |
33 ~ 40 VDC |
3.63 ~ 3.00 A |
2% |
120W |
Â
Electrical Characteristics |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Safety Approvals Input Voltage Range |
 |
100 |
 |
240 |
VAC |
Input Voltage |
Operating Voltage |
90 |
 |
260 |
VAC |
Input Frequency |
 |
47 |
 |
63 |
Hz |
Power Factor Correction |
Io=Full load, Vin=90~260VAC |
0.95 |
 |
1.00 |
 |
Output Power Range |
Vin= 90 to 264 VA C |
See rating chart |
W |
||
Input Current (Low Line) |
Io=Full load, Vin=100 VAC |
 |
 |
1.7 |
A |
Input Current (High Line) |
Io=Full load, Vin= 240 VAC |
 |
 |
0.8 |
A |
Low Line Inrush Current |
Io=Full load, 25°C, Cool start, Vin=115VAC |
 |
 |
30 |
A |
High Line Inrush Current |
Io=Full load, 25°C, Cool start, Vin=230VAC |
 |
 |
60 |
A |
Efficiency |
Io=Full Load, Vin=230VAC |
72.5 |
 |
88 |
% |
Line Regulation |
Io=Full Load |
 |
 |
1 |
% |
Load Regulation |
Vin=230VAC |
 |
 |
5 |
% |
Over Voltage Protection |
 |
112 |
 |
132 |
% |
Over Current Protection |
 |
110 |
 |
150 |
% |
Transient Response |
Io=Full Load to Half Load, Vin=100VAC |
 |
 |
4 |
mS |
Hold-Up Time |
Io=Full Load, Vin=110VAC |
16 |
 |
 |
mS |
Start Up Time |
Io=Full Load, Vin=100VAC |
0.3 |
 |
2 |
S |
Ripple & Noise (Peak to Peak) |
Full Load, Vin=90VAC |
 |
 |
1 |
% |
Safety Ground Leakage Current |
Io= Full Load, Vin=240VAC/60Hz |
 |
 |
0.1 |
mA |
Temperature Coefficient |
All output |
-0.04 |
 |
0.04 |
%/°C |
Â
Conditions |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Operating Temperature |
 |
-10 |
 |
70 |
°C |
Storage Temperature |
 |
-40 |
 |
85 |
°C |
Operating Humidity |
Non-condensed |
0 |
 |
95 |
% |
Storage Humidity |
Non-condensed |
0 |
 |
95 |
% |
Operating Temperature at 25°C, Calculated per MIL-HDBK-217F |
100K |
 |
 |
 Hrs |
|
De-rate linearly from 100% load at 50°C to 50% load at 70°C |
Â
Approvals and Compliance |
|||
Parameter |
Test Conditions |
Min. |
Unit |
Dielectric Withstanding Voltage for Primary to secondary |
Primary to secondary |
4000 |
VAC |
Dielectric Withstanding Voltage for Primary to PE |
Primary to PE |
1500 |
VAC |
Isolation Resistance |
Test Voltage = 500VDC |
50 |
MΩ |
EMI (EMC Emission) |
Compliance to EN55011 (CISPR11), EN60601-1-2 |
B |
CLASS |
Safety UL/c-UL, TUV/T-mark, CE, FCC |
 UL/c-UL(UL 60601-1:2ndEdition), TUV/T-mark(IEC 60601-1:2 Safety ndEdition) |
Â
Mechanical and PIN out |
   Â
PIN CHART
Â
|
Note:
|
||||||||||||||||||||||||||||||||||||||||
 |
Â