Specifications
3 Year Warranty |
|
Approvals: |
Single Output |
Product No. |
Setting Voltage Range (Factory setting, can‘t be adjusted) |
Output Current (Based on the output volt.) |
Max Output Power |
Typ. Ripple & Noise |
Total Regulation |
Typ. Efficiency |
Typ. |
Hold-Up Time |
Protection Mode |
||
min |
max |
min |
max |
||||||||
(VDC) |
(VDC) |
(A) |
(A) |
(W) |
(mVp-p) |
(%) |
(%) |
(W) |
(ms) |
||
SHFA32-102 |
5.0 |
6.0 |
3.33 |
4.00 |
20 |
50 |
±5 |
74 |
0.5 |
12 |
Hiccup |
SHFA32-S03 |
6.0 |
8.0 |
2.87 |
3.83 |
23 |
70 |
±5 |
77 |
0.5 |
12 |
Hiccup |
SHFA32-S04 |
8.0 |
11.0 |
2.45 |
3.37 |
27 |
90 |
±5 |
78 |
0.5 |
12 |
Hiccup |
SHFA32-S05 |
11.0 |
13.0 |
2.31 |
2.72 |
30 |
100 |
±5 |
83 |
0.5 |
12 |
Hiccup |
SHFA32-S06 |
13.0 |
16.0 |
1.88 |
2.31 |
30 |
100 |
±5 |
84 |
0.5 |
12 |
Hiccup |
SHFA32-S07 |
16.0 |
21.0 |
1.43 |
1.88 |
30 |
100 |
±5 |
85 |
0.5 |
12 |
Hiccup |
SHFA32-S08 |
21.0 |
27.0 |
1.12 |
1.43 |
30 |
100 |
±3 |
86 |
0.5 |
12 |
Hiccup |
SHFA32-S09 |
27.0 |
33.0 |
0.91 |
1.12 |
30 |
100 |
±3 |
86 |
0.5 |
12 |
Hiccup |
SHFA32-S10 |
33.0 |
40.0 |
0.76 |
0.91 |
30 |
100 |
±3 |
86 |
0.5 |
12 |
Hiccup |
Electrical Characteristics |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Safety Approvals Input Voltage Range |
Safety Approval & Specification in Label |
100 |
|
240 |
VAC |
Operate Voltage Range |
Derate linearly from 100% load at 90VAC to 80% load at 80VAC |
80 |
|
275 |
VAC |
Input Frequency |
Sine wave |
47 |
|
63 |
Hz |
Output Power Range |
See Rating Chart |
|
|
30 |
W |
Low Line Input Current |
Full load, Vin=100VAC |
|
|
0.8 |
A |
High Line Input Current |
Full load, Vin=240VAC |
|
|
0.4 |
A |
Low Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=100VAC |
|
|
25 |
A |
High Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=240VAC |
|
|
50 |
A |
Efficiency |
Full Load, Vin=230VAC |
74 |
|
86 |
% |
Line Regulation |
Full Load, Vin=100~120VAC or 200~240VAC |
|
|
1 |
% |
Over Load Protection |
Recovers automatically after fault condition is removed |
110 |
|
150 |
% |
Time of Transient Response |
Io=Full Load to Half Load, Vin=110VAC |
|
|
4 |
ms |
Hold-Up Time |
Full Load, Vin=110VAC |
|
|
12 |
ms |
Start Up Time |
Full Load, Vin=100~240VAC |
|
|
0.5 |
s |
Insulation Resistance |
|
50 |
|
|
MΩ |
Ripple & Noise (Peak to Peak) |
|
|
|
1 |
% |
Temperature Coefficient |
All output |
|
|
±0.04 |
%/°C |
Dielectric Withstanding Voltage(P-S) |
Primary to Secondary, limit current<10mA |
|
|
4000 |
VAC |
EMC Emission |
Compliance to EN55011(CISPR11), EN60601-1-2 |
B |
|
|
Class |
Environmental |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Operating Temperature |
Derate linearly from 100% load at 40℃ to 50% load at 70℃ |
-10 |
|
70 |
°C |
Storage Temperature |
10~95% RH |
-40 |
|
85 |
°C |
Operating Humidity |
non-condensing |
0 |
|
95% |
RH |
Storage Humidity |
|
0 |
|
95% |
RH |
Electro Static Discharge |
Air Discharge, IEC61000-4-2 |
|
|
15 |
KV |
Electro Static Discharge |
Contact Discharge, IEC61000-4-2 |
|
|
8 |
KV |
Mean Time Between Failure |
Operation Temperature at 25℃J, Calculated per MIL-HDBK-217F |
200K |
|
|
h |
Operating Altitude (Elevation) |
All Condition |
|
|
3000 |
m |
Vibration |
10~500Hz,10min./1cycle, 60min.each along X, Y, Z axes |
|
|
5 |
G |
Surge Voltage |
Line-Neutral |
|
|
1 |
KV |
Surge Voltage |
Line-PE & Neutral-PE |
|
|
2 |
KV |
Mechanical Drawing & Technical Chart |
PIN Chart:
Note:
|
Derating Curve |
Datasheet PDF