Specifications
Â
3 Year Warranty |
 |
Approvals:Â Â |
Â
Single Output |
||||
Model Number |
Output Voltage |
Max. Output Current |
Total Regulation |
Maximum Output Power |
SHDA180B -S05 |
12 VDC |
14 A |
±5% |
168W |
SHDA180B -S07 |
19 VDC |
9.47 A |
±5% |
180W |
SHDA180B -S08 |
24 VDC |
7.50 A |
±4% |
180W |
SHDA180B -S09 |
30 VDC |
6.00 A |
±3% |
180W |
SHDA180B -S10 |
33 VDC |
5.455 A |
±3% |
180W |
SHDA180B -S11 |
48 VDC |
3.75 A |
±3% |
180W |
Selected output connectors and wire, please refer to Appendix.
This series is required  to use AWG#16/5C/4FT output cable.
The regulation and efficiency  will be changed by modified output cable.
Â
Electrical Characteristics |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Safety Approvals Input Voltage Range |
Safety Approval & Specification in Label |
100 |
 |
240 |
VAC |
Operate Voltage Range |
 |
90 |
 |
260 |
VAC |
Input Frequency |
Sine wave |
47 |
 |
63 |
Hz |
Power Factor Correction |
 |
0.95 |
 |
1 |
 |
Output Power Range |
See Rating Chart |
 |
 |
180 |
W |
Low Line Input Current |
Full load, Vin=100VAC |
 |
 |
2.2 |
A |
High Line Input Current |
Full load, Vin=240VAC |
 |
 |
0.9 |
A |
Low Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=100VAC |
 |
 |
60 |
A |
High Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=240VAC |
 |
 |
120 |
A |
Safety Ground Leakage Current |
Vin=240VAC, Fi=60Hz |
 |
 |
0.25 |
mA |
Efficiency |
Full Load, Vin=230VAC |
89 |
 |
93 |
% |
Line Regulation |
Full Load, Vin=100~120VAC or 200~240VAC |
 |
 |
1 |
% |
Over Voltage Protection |
 |
112 |
 |
132 |
% |
Over Load Protection |
Recovers automatically after fault condition is removed |
110 |
 |
150 |
% |
Time of Transient Response |
Io=Full Load to Half Load, Vin=110VAC |
 |
 |
4 |
ms |
Hold-Up Time |
Full Load, Vin=110VAC |
 |
 |
20 |
ms |
Start Up Time |
Full Load, Vin=100~240VAC |
 |
 |
2 |
s |
Ripple & Noise (Peak to Peak) |
 |
 |
 |
1 |
% |
Temperature Coefficient |
All output |
 |
 |
±0.04 |
%/°C |
Dielectric Withstanding Voltage(P-S) |
Primary to Secondary, limit current<10mA |
 |
 |
4000 |
VAC |
EMC Emission |
Compliance to EN55011(CISPR11), EN60601-1-2 |
B |
 |
 |
Class |
Â
Â
Environmental |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Operating Temperature |
Derate linearly from 100% load at 40â to 50% load at 70â |
-10 |
 |
70 |
°C |
Storage Temperature |
10~95% RH |
-40 |
 |
85 |
°C |
Operating Humidity |
non-condensing |
0 |
 |
95% |
RH |
Storage Humidity |
 |
0 |
 |
95% |
RH |
Electro Static Discharge |
Air Discharge, IEC61000-4-2 |
 |
 |
15 |
KV |
Electro Static Discharge |
Contact Discharge, IEC61000-4-2 |
 |
 |
8 |
KV |
Mean Time Between Failure |
Operation Temperature at 25âJ, Calculated per MIL-HDBK-217F |
200K |
 |
 |
h |
Operating Altitude (Elevation) |
All Condition |
 |
 |
3000 |
m |
Vibration |
10~500Hz,10min./1cycle, 60min.each along X, Y, Z axes |
 |
 |
5 |
G |
Surge Voltage |
 Line-Neutral |
 |
 |
1 |
KV |
Surge Voltage |
 Line-PE & Neutral-PE |
 |
 |
2 |
KV |
Â
Mechanical Diagram and Technical Chart |
Note: Â
|
Output Derating Curve
|
Â
Datasheet PDF