Specifications
Â
3 Year Warranty |
 |
Approvals:Â |
Â
Single Output |
|||||
Part Number |
Output Voltage |
Max. Output Current |
Total Regulation |
Max. Output Power |
Pno (max.) |
SHDA101-S05 |
12 ~ 13 VDC |
8.33 ~ 7.70 A |
5% |
100W |
0.5W |
SHDA101-S06 |
13 ~ 16 VDC |
7.70 ~ 6.30 A |
4% |
100W |
0.5W |
SHDA101-S07 |
16 ~ 21 VDC |
6.30 ~ 4.80 A |
4% |
100W |
0.5W |
SHDA101-S08 |
21 ~ 27 VDC |
4.80 ~ 3.70 A |
4% |
100W |
0.5W |
SHDA101-S09 |
27 ~ 33 VDC |
3.70 ~ 3.00 A |
3% |
100W |
0.5W |
SHDA101-S10 |
33 ~ 40 VDC |
3.00 ~ 2.50 A |
3% |
100W |
0.5W |
SHDA101-S11 |
40 ~ 48 VDC |
2.50 ~ 2.08 A |
3% |
100W |
0.5W |
 The total regulation each model is required to use AWG#18/3C+AWG16/3C/4FT output cable
The regulation will be changed by modified output cable.
Â
Electrical Characteristics |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Safety Approvals Input Voltage Range |
Safety Approval & Specification in Label |
100 |
 |
240 |
VAC |
Operate Voltage Range |
 |
90 |
 |
260 |
VAC |
Input Frequency |
Sine wave |
47 |
 |
63 |
Hz |
Power Factor Correction |
 |
0.95 |
 |
1 |
 |
Output Power Range |
See Rating Chart |
 |
 |
100 |
W |
Low Line Input Current |
Full load, Vin=100VAC |
 |
 |
1.2 |
A |
High Line Input Current |
Full load, Vin=240VAC |
 |
 |
0.5 |
A |
Low Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=100VAC |
 |
 |
50 |
A |
High Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=240VAC |
 |
 |
120 |
A |
Safety Ground Leakage Current |
Vin=240VAC, Fi=60Hz |
 |
 |
0.25 |
mA |
Efficiency |
Full Load, Vin=230VAC |
81 |
 |
87 |
% |
Line Regulation |
Full Load, Vin=100~120VAC or 200~240VAC |
0.5 |
 |
1 |
% |
Over Voltage Protection |
 |
112 |
 |
132 |
% |
Over Load Protection |
Recovers automatically after fault condition is removed |
110 |
 |
150 |
% |
Time of Transient Response |
Io=Full Load to Half Load, Vin=110VAC |
 |
 |
4 |
ms |
Hold-Up Time |
Full Load, Vin=100VAC |
 |
 |
16 |
ms |
Start Up Time |
Full Load, Vin=100~240VAC |
0.3 |
 |
2 |
s |
Insulation Resistance |
 |
50 |
 |
 |
MΩ |
Ripple & Noise (Peak to Peak) |
 |
 |
 |
1 |
% |
Temperature Coefficient |
All output |
 |
 |
±0.04 |
%/°C |
Dielectric Withstanding Voltage(P-S) |
Primary to Secondary, limit current<10mA |
 |
 |
4000 |
VAC |
Dielectric Withstanding Voltage(P-G) |
Primary to PE, limit current<10mA |
 |
 |
1500 |
VAC |
EMC Emission |
Compliance to EN55011(CISPR11), EN60601-1-2 |
B |
 |
 |
Class |
Â
Â
Environmental |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Operating Temperature |
Derate linearly from 100% load at 40â to 50% load at 70â |
-10 |
 |
70 |
°C |
Storage Temperature |
10~95% RH |
-40 |
 |
85 |
°C |
Operating Humidity |
non-condensing |
0 |
 |
95% |
RH |
Storage Humidity |
 |
0 |
 |
95% |
RH |
Electro Static Discharge |
Air Discharge, IEC61000-4-2 |
 |
 |
15 |
KV |
Electro Static Discharge |
Contact Discharge, IEC61000-4-2 |
 |
 |
8 |
KV |
Mean Time Between Failure |
Operation Temperature at 25âJ, Calculated per MIL-HDBK-217F |
200K |
 |
 |
h |
Operating Altitude (Elevation) |
All Condition |
 |
 |
3000 |
m |
Vibration |
10~500Hz,10min./1cycle, 60min.each along X, Y, Z axes |
 |
 |
5 |
G |
Surge Voltage |
Line-Neutral |
 |
 |
1 |
KV |
Surge Voltage |
Line-PE & Neutral-PE |
 |
 |
2 |
KV |
Â
Â
Mechanical Diagram and Derating Chart |
Â
Note:
(Exclude the input cord)       Optional output connector.    Derating Chart  1. Operating Temperature: -10 to 70°C 2. Derate linearly from 100% load at 50°C to 50% load at 70°C |
Â
Â
Â
Datasheet PDF
Â
Â