SHDA101 SERIES


100W Desktop Power Supply for Medical Equipment


  • Wide Operating Voltage, 90 to 260 VAC, 47 to 63 Hz
  • IEC-320-C14 Input Inlet
  • Single Output
  • Crowbar Mode Over Voltage Protection
  • Input to Output : 2MOPP
  • High ESD immunity
  • Suitable professional healthcare facility
  • Low earth leakage current < 0.25mA

Specifications

 

 

3 Year Warranty

 

Approvals: 

 

Single Output

Part Number

Output Voltage

Max. Output Current

Total Regulation

Max. Output Power

Pno (max.)

SHDA101-S05

12 ~ 13 VDC

8.33 ~ 7.70 A

5%

100W

0.5W

SHDA101-S06

13 ~ 16 VDC

7.70 ~ 6.30 A

4%

100W

0.5W

SHDA101-S07

16 ~ 21 VDC

6.30 ~ 4.80 A

4%

100W

0.5W

SHDA101-S08

21 ~ 27 VDC

4.80 ~ 3.70 A

4%

100W

0.5W

SHDA101-S09

27 ~ 33 VDC

3.70 ~ 3.00 A

3%

100W

0.5W

SHDA101-S10

33 ~ 40 VDC

3.00 ~ 2.50 A

3%

100W

0.5W

SHDA101-S11

40 ~ 48 VDC

2.50 ~ 2.08 A

3%

100W

0.5W

 The total regulation each model is required to use AWG#18/3C+AWG16/3C/4FT output cable

The regulation will be changed by modified output cable.

 

Electrical Characteristics

Parameter

Test Conditions

Min.

Typ.

Max.

Unit

Safety Approvals Input Voltage Range

Safety Approval & Specification in Label

100

 

240

VAC

Operate Voltage Range

 

90

 

260

VAC

Input Frequency

Sine wave

47

 

63

Hz

Power Factor Correction

 

0.95

 

1

 

Output Power Range

See Rating Chart

 

 

100

W

Low Line Input Current

Full load, Vin=100VAC

 

 

1.2

A

High Line Input Current

Full load, Vin=240VAC

 

 

0.5

A

Low Line Input Inrush Current

Full load, 25°C, Cool start, Vin=100VAC

 

 

50

A

High Line Input Inrush Current

Full load, 25°C, Cool start, Vin=240VAC

 

 

120

A

Safety Ground Leakage Current

Vin=240VAC, Fi=60Hz

 

 

0.25

mA

Efficiency

Full Load, Vin=230VAC

81

 

87

%

Line Regulation

Full Load, Vin=100~120VAC or 200~240VAC

0.5

 

1

%

Over Voltage Protection

 

112

 

132

%

Over Load Protection

Recovers automatically after fault condition is removed

110

 

150

%

Time of Transient Response

Io=Full Load to Half Load, Vin=110VAC

 

 

4

ms

Hold-Up Time

Full Load, Vin=100VAC

 

 

16

ms

Start Up Time

Full Load, Vin=100~240VAC

0.3

 

2

s

Insulation Resistance

 

50

 

 

MΩ

Ripple & Noise (Peak to Peak)

 

 

 

1

%

Temperature Coefficient

All output

 

 

±0.04

%/°C

Dielectric Withstanding Voltage(P-S)

Primary to Secondary, limit current<10mA

 

 

4000

VAC

Dielectric Withstanding Voltage(P-G)

Primary to PE, limit current<10mA

 

 

1500

VAC

EMC Emission

Compliance to EN55011(CISPR11), EN60601-1-2

B

 

 

Class

 

 

Environmental

Parameter

Test Conditions

Min.

Typ.

Max.

Unit

Operating Temperature

Derate linearly from 100% load at 40℃ to 50% load at 70℃

-10

 

70

°C

Storage Temperature

10~95% RH

-40

 

85

°C

Operating Humidity

non-condensing

0

 

95%

RH

Storage Humidity

 

0

 

95%

RH

Electro Static Discharge

Air Discharge, IEC61000-4-2

 

 

15

KV

Electro Static Discharge

Contact Discharge, IEC61000-4-2

 

 

8

KV

Mean Time Between Failure

Operation Temperature at 25℃J, Calculated per MIL-HDBK-217F

200K

 

 

h

Operating Altitude (Elevation)

All Condition

 

 

3000

m

Vibration

10~500Hz,10min./1cycle, 60min.each along X, Y, Z axes

 

 

5

G

Surge Voltage

Line-Neutral

 

 

1

KV

Surge Voltage

Line-PE & Neutral-PE

 

 

2

KV

 

 

Mechanical Diagram and Derating Chart

 

Note:

  1. Dimensions are shown in mm & inch
  2. Weight: approx. 490-670gs

(Exclude the input cord)

           Optional output connector.

 

 

 

Derating Chart

 

1. Operating Temperature: -10 to 70°C

2. Derate linearly from 100% load at 50°C to 50% load at 70°C

 

 

 

SHDA101 - MEPOS ELECTRONICS