Specifications
3 Year Warranty |
|
Approvals: |
Single Output |
||||
Model Number |
Output Voltage |
Max. Output Current |
Load Regulation |
Maximum Output Power |
SMDA32B-S02 |
5 ~ 5.99 VDC |
4.00 ~ 3.33 A |
±5% |
20W |
SMDA32B-S03 |
6 ~ 8 VDC |
3.83 ~ 2.87 A |
±5% |
23W |
SMDA32B-S04 |
8 ~ 11 VDC |
3.38 ~ 2.45 A |
±5% |
27W |
SMDA32B-S05 |
11 ~ 13 VDC |
2.74 ~ 2.30 A |
±5% |
30W |
SMDA32B-S06 |
13 ~ 16 VDC |
2.30 ~ 1.88 A |
±5% |
30W |
SMDA32B-S07 |
16 ~ 21 VDC |
1.88 ~ 1.43 A |
±5% |
30W |
SMDA32B-S08 |
21 ~ 27 VDC |
1.43 ~ 1.11 A |
±3% |
30W |
SMDA32B-S09 |
27 ~ 33 VDC |
1.11 ~ 0.91A |
±3% |
30W |
SMDA32B-S10 |
33 ~ 40 VDC |
0.91 ~ 0.75A |
±3% |
30W |
SMDA32B-S11 |
40 ~ 48 VDC |
0.75 ~ 0.63A |
±3% |
30W |
Total Regulation is conditioned by below configuration
(SHDA32B-S02~S04: AWG16/4FT output cable)
(SHDA32B-S05~S11: AWG18/6FT output cable)
Electrical Characteristics |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Safety Approvals Input Voltage Range |
Safety Approval & Specification in Label |
100 |
|
240 |
VAC |
Operate Voltage Range |
Derate linearly from 100% load at 90VAC to 80% load at 80VAC |
80 |
|
275 |
VAC |
Input Frequency |
Sine wave |
47 |
|
63 |
Hz |
Output Power Range |
See Rating Chart |
|
|
30 |
W |
Low Line Input Current |
Full load, Vin=100VAC |
|
|
0.6 |
A |
High Line Input Current |
Full load, Vin=240VAC |
|
|
0.4 |
A |
Low Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=100VAC |
|
|
28 |
A |
High Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=240VAC |
|
|
67 |
A |
Efficiency |
Full Load, Vin=230VAC |
80 |
|
86 |
% |
Line Regulation |
Full Load, Vin=100~120VAC |
|
|
1 |
% |
Over Voltage Protection |
|
112 |
|
132 |
% |
Over Load Protection |
Recovers automatically after fault condition is removed |
110 |
|
150 |
% |
Time of Transient Response |
Io=Full Load to Half Load, Vin=110VAC |
|
|
4 |
ms |
Hold-Up Time |
Full Load, Vin=100VAC |
|
|
12 |
ms |
Start Up Time |
Full Load, Vin=100~240VAC |
|
|
2 |
s |
Ripple & Noise (Peak to Peak) |
|
|
|
1 |
% |
Temperature Coefficient |
All output |
|
|
±0.04 |
%/°C |
Dielectric Withstanding Voltage(P-S) |
Primary to Secondary, limit current<10mA |
|
|
4000 |
VAC |
EMC Emission |
Compliance to EN55011(CISPR11), EN60601-1-2 |
B |
|
|
Class |
Environmental |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Operating Temperature |
Derate linearly from 100% load at 50℃ to 50% load at 70℃ |
-10 |
|
70 |
°C |
Storage Temperature |
10~95% RH |
-40 |
|
85 |
°C |
Operating Humidity |
non-condensing |
0 |
|
95% |
RH |
Storage Humidity |
|
0 |
|
95% |
RH |
Electro Static Discharge |
Air Discharge, IEC61000-4-2 |
|
|
15 |
KV |
Electro Static Discharge |
Contact Discharge, IEC61000-4-2 |
|
|
8 |
KV |
Mean Time Between Failure |
Operation Temperature at 25℃J, Calculated per MIL-HDBK-217F |
200K |
|
|
h |
Operating Altitude (Elevation) |
All Condition |
|
|
5000 |
m |
Vibration |
10~500Hz,10min./1cycle, 60min.each along X, Y, Z axes |
|
|
5 |
G |
Surge Voltage |
Line-Neutral |
|
|
1 |
KV |
Surge Voltage |
Line-PE & Neutral-PE |
|
|
2 |
KV |
Mechanical Diagram |
Derating Curve |
Datasheet PDF