Specifications
Â
3 Year Warranty |
 |
Approvals:Â |
Â
Single Output |
||||
Model Number |
Output Voltage |
Max. Output Current |
Load Regulation |
Maximum Output Power |
SMDA32A-S02Â |
5 ~ 5.99 VDC |
4.00 ~ 3.33 A |
±5% |
20W |
SMDA32A-S03Â |
6 ~ 8 VDC |
3.83 ~ 2.87 A |
±5% |
23W |
SMDA32A-S04 |
8 ~ 11 VDC |
3.38 ~ 2.45 A |
±5% |
27W |
SMDA32A-S05 |
11 ~ 13 VDC |
2.74 ~ 2.30 A |
±5% |
30W |
SMDA32A-S06 |
13 ~ 16 VDC |
2.30 ~ 1.88 A |
±5% |
30W |
SMDA32A-S07 |
16 ~ 21 VDC |
1.88 ~ 1.43 A |
±5% |
30W |
SMDA32A-S08 |
21 ~ 27 VDC |
1.43 ~ 1.11 A |
±3% |
30W |
SMDA32A-S09 |
27 ~ 33 VDC |
1.11 ~ 0.91A |
±3% |
30W |
SMDA32A-S10 |
33 ~ 40 VDC |
0.91 ~ 0.75A |
±3% |
30W |
SMDA32A-S11 |
40 ~ 48 VDC |
0.75 ~ 0.63A |
±3% |
30W |
Total Regulation is conditioned by below configuration
(SHDA32A-S02~S04: AWG16/4FT output cable)
(SHDA32B-S05~S11: AWG18/6FT output cable)
Â
Electrical Characteristics |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Safety Approvals Input Voltage Range |
Safety Approval & Specification in Label |
100 |
 |
240 |
VAC |
Operate Voltage Range |
Derate linearly from 100% load at 90VAC to 80% load at 80VAC |
80 |
 |
275 |
VAC |
Input Frequency |
Sine wave |
47 |
 |
63 |
Hz |
Output Power Range |
See Rating Chart |
 |
 |
30 |
W |
Low Line Input Current |
Full load, Vin=100VAC |
 |
 |
0.6 |
A |
High Line Input Current |
Full load, Vin=240VAC |
 |
 |
0.4 |
A |
Low Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=100VAC |
 |
 |
28 |
A |
High Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=240VAC |
 |
 |
67 |
A |
Efficiency |
Full Load, Vin=230VAC |
80 |
 |
86 |
% |
Line Regulation |
Full Load, Vin=100~120VAC |
 |
 |
1 |
% |
Over Voltage Protection |
 |
112 |
 |
132 |
% |
Over Load Protection |
Recovers automatically after fault condition is removed |
110 |
 |
150 |
% |
Time of Transient Response |
Io=Full Load to Half Load, Vin=110VAC |
 |
 |
4 |
ms |
Hold-Up Time |
Full Load, Vin=100VAC |
 |
 |
12 |
ms |
Start Up Time |
Full Load, Vin=100~240VAC |
 |
 |
2 |
s |
Ripple & Noise (Peak to Peak) |
 |
 |
 |
1 |
% |
Temperature Coefficient |
All output |
 |
 |
±0.04 |
%/°C |
Dielectric Withstanding Voltage(P-S) |
Primary to Secondary, limit current<10mA |
 |
 |
4000 |
VAC |
EMC Emission |
Compliance to EN55011(CISPR11), EN60601-1-2 |
B |
 |
 |
Class |
Â
Â
Environmental |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Operating Temperature |
Derate linearly from 100% load at 50â to 50% load at 70â |
-10 |
 |
70 |
°C |
Storage Temperature |
10~95% RH |
-40 |
 |
85 |
°C |
Operating Humidity |
non-condensing |
0 |
 |
95% |
RH |
Storage Humidity |
 |
0 |
 |
95% |
RH |
Electro Static Discharge |
Air Discharge, IEC61000-4-2 |
 |
 |
15 |
KV |
Electro Static Discharge |
Contact Discharge, IEC61000-4-2 |
 |
 |
8 |
KV |
Mean Time Between Failure |
Operation Temperature at 25âJ, Calculated per MIL-HDBK-217F |
200K |
 |
 |
h |
Operating Altitude (Elevation) |
All Condition |
 |
 |
5000 |
m |
Vibration |
10~500Hz,10min./1cycle, 60min.each along X, Y, Z axes |
 |
 |
5 |
G |
Surge Voltage |
 Line-Neutral |
 |
 |
1 |
KV |
Surge Voltage |
 Line-PE & Neutral-PE |
 |
 |
2 |
KV |
Â
Mechanical Diagram |
Â
Â
Derating Curve |
1. Operating Temperature: -10 to 70°C 2. Derate linearly from 100% load at 50°C to 50% load at 70°C |
Â
Â