Print this page

SHWA25 SERIES


25W Multi-Plug Wall Mount for Medical Equipment


  • Wide Operating Voltage 80 to 275VAC, 47 to 63Hz
  • Interchangeable AC plug(USA, UK, Europe, Australia)
  • Single Output
  • Input to Output : 2MOPP
  • High ESD immunity
  • Suitable home healthcare environment
  • Suitable professional healthcare facility

Specifications

 

 

3 Year Warranty

 

Approvals:

 

Single Output

Model Number

Output Voltage

Max. Output Current

Total Regulation

Max. Output Power

SHWA25-S01

3 ~ 5 VDC

2.10 ~ 3.50 A

±5%

10.5W

SHWA25-S02

5 ~ 6 VDC

2.75 ~ 3.30 A

±5%

16.5W

SHWA25-S03

6 ~ 8 VDC

2.50 ~ 3.30 A

±5%

20W

SHWA25-S04

8 ~ 11 VDC

2.00 ~ 2.75 A

±5%

22W

SHWA25-S05

11 ~ 13 VDC

1.92 ~ 2.27 A

±5%

25W

SHWA25-S06

13 ~ 16 VDC

1.56 ~ 1.92 A

±5%

25W

SHWA25-S07

16 ~ 21 VDC

1.19 ~ 1.56 A

±5%

25W

SHWA25-S08

21 ~ 27 VDC

0.92 ~ 1.19 A

±3%

25W

SHWA25-S09

27 ~ 33 VDC

0.75 ~ 0.92 A

±3%

25W

SHWA25-S10

33 ~ 40 VDC

0.62 ~ 0.75 A

±3%

25W

SHWA25-S11

40 ~ 48 VDC

0.53 ~ 0.62 A

±3%

25W

SHWA25-S12

48 ~ 55 VDC

0.45 ~ 0.53 A

±3%

25W

SHWA25-S01~S04 required to use AWG#18—2C/4FT output cable.

SHWA25-S05~S09 required to use AWG#20—2C/4FT output cable.

SHWA25-S10~S12 required to use AWG#22—2C/4FT output cable.

The regulation and efficiency will be changed by modified output cable.

 

Electrical Characteristics

Parameter

Test Conditions

Min.

Typ.

Max.

Unit

Safety Approvals Input Voltage Range

Safety Approval & Specification in Label

100

 

240

VAC

Operate Voltage Range

Derate linearly from 100% load at 90VAC to 80% load at 80VAC

80

 

275

VAC

Input Frequency

Sine wave

47

 

63

Hz

Output Power Range

See Rating Chart

 

 

25

W

Low Line Input Current

Full load, Vin=100VAC

 

 

0.70

A

High Line Input Current

Full load, Vin=240VAC

 

 

0.40

A

Low Line Input Inrush Current

Full load, 25°C, Cool start, Vin=100VAC

 

 

65

A

High Line Input Inrush Current

Full load, 25°C, Cool start, Vin=240VAC

 

 

165

A

Efficiency

Full Load, Vin=230VAC

65

 

86

%

Line Regulation

Full Load, Vin=100~120VAC

 

 

1

%

Over Load Protection

Recovers automatically after fault condition is removed

110

 

150

%

Time of Transient Response

Io=Full Load to Half Load, Vin=110VAC

 

 

4

ms

Hold-Up Time

Full Load, Vin=100VAC

 

 

12

ms

Start Up Time

Full Load, Vin=100~240VAC

 

 

3

s

* Ripple & Noise (Peak to Peak)

 

 

1

2

%

Temperature Coefficient

All output

 

 

±0.04

%/°C

Dielectric Withstanding Voltage(P-S)

Primary to Secondary, limit current<10mA

 

 

4000

VAC

EMC Emission

Compliance to EN55011(CISPR11), EN60601-1-2

B

 

 

Class

 

 

Environmental

Parameter

Test Conditions

Min.

Typ.

Max.

Unit

Operating Temperature

Derate linearly from 100% load at 40°C to 50% load at 70°C

-10

 

70

°C

Storage Temperature

  10~95% RH

-40

 

85

°C

Operating Humidity

  non-condensing

0

 

95%

RH

Storage Humidity

 

0

 

95%

RH

Electro Static Discharge

Air Discharge, IEC61000-4-2

 

 

15

KV

Electro Static Discharge

Contact Discharge, IEC61000-4-2

 

 

8

KV

Mean Time Between Failure

Operation Temperature at 25°C, Calculated per MIL-HDBK-217F

200K

 

 

h

Operating Altitude (Elevation)

All Condition

 

 

3000

m

Vibration

10~500Hz,10min./1cycle, 60min.each along X, Y, Z axes

 

 

5

G

Surge Voltage

  Line-Neutral

 

 

1

KV

Surge Voltage

  Line-PE & Neutral-PE

 

 

2

KV

 

Mechanical Drawing and Technical Chart

      

Note:

  1. Dimensions are shown in mm.
  2. Weight: 200g approx.  
  3. Optional output connector.

Output Derating Chart:

1. Operating Temperature: -10 to 70°C

2. Derate linearly from 100% load at 40°C to 50% load at 70°C

 

Datasheet PDF