Specifications
3 Year Warranty |
|
Approvals: |
Single Output |
||||
Model Number |
Output Voltage |
Max. Output Current |
Total Regulation |
Max. Output Power |
SHWA25-S01 |
3 ~ 5 VDC |
2.10 ~ 3.50 A |
±5% |
10.5W |
SHWA25-S02 |
5 ~ 6 VDC |
2.75 ~ 3.30 A |
±5% |
16.5W |
SHWA25-S03 |
6 ~ 8 VDC |
2.50 ~ 3.30 A |
±5% |
20W |
SHWA25-S04 |
8 ~ 11 VDC |
2.00 ~ 2.75 A |
±5% |
22W |
SHWA25-S05 |
11 ~ 13 VDC |
1.92 ~ 2.27 A |
±5% |
25W |
SHWA25-S06 |
13 ~ 16 VDC |
1.56 ~ 1.92 A |
±5% |
25W |
SHWA25-S07 |
16 ~ 21 VDC |
1.19 ~ 1.56 A |
±5% |
25W |
SHWA25-S08 |
21 ~ 27 VDC |
0.92 ~ 1.19 A |
±3% |
25W |
SHWA25-S09 |
27 ~ 33 VDC |
0.75 ~ 0.92 A |
±3% |
25W |
SHWA25-S10 |
33 ~ 40 VDC |
0.62 ~ 0.75 A |
±3% |
25W |
SHWA25-S11 |
40 ~ 48 VDC |
0.53 ~ 0.62 A |
±3% |
25W |
SHWA25-S12 |
48 ~ 55 VDC |
0.45 ~ 0.53 A |
±3% |
25W |
SHWA25-S01~S04 required to use AWG#182C/4FT output cable.
SHWA25-S05~S09 required to use AWG#202C/4FT output cable.
SHWA25-S10~S12 required to use AWG#222C/4FT output cable.
The regulation and efficiency will be changed by modified output cable.
Electrical Characteristics |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Safety Approvals Input Voltage Range |
Safety Approval & Specification in Label |
100 |
|
240 |
VAC |
Operate Voltage Range |
Derate linearly from 100% load at 90VAC to 80% load at 80VAC |
80 |
|
275 |
VAC |
Input Frequency |
Sine wave |
47 |
|
63 |
Hz |
Output Power Range |
See Rating Chart |
|
|
25 |
W |
Low Line Input Current |
Full load, Vin=100VAC |
|
|
0.70 |
A |
High Line Input Current |
Full load, Vin=240VAC |
|
|
0.40 |
A |
Low Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=100VAC |
|
|
65 |
A |
High Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=240VAC |
|
|
165 |
A |
Efficiency |
Full Load, Vin=230VAC |
65 |
|
86 |
% |
Line Regulation |
Full Load, Vin=100~120VAC |
|
|
1 |
% |
Over Load Protection |
Recovers automatically after fault condition is removed |
110 |
|
150 |
% |
Time of Transient Response |
Io=Full Load to Half Load, Vin=110VAC |
|
|
4 |
ms |
Hold-Up Time |
Full Load, Vin=100VAC |
|
|
12 |
ms |
Start Up Time |
Full Load, Vin=100~240VAC |
|
|
3 |
s |
* Ripple & Noise (Peak to Peak) |
|
|
1 |
2 |
% |
Temperature Coefficient |
All output |
|
|
±0.04 |
%/°C |
Dielectric Withstanding Voltage(P-S) |
Primary to Secondary, limit current<10mA |
|
|
4000 |
VAC |
EMC Emission |
Compliance to EN55011(CISPR11), EN60601-1-2 |
B |
|
|
Class |
Environmental |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Operating Temperature |
Derate linearly from 100% load at 40°C to 50% load at 70°C |
-10 |
|
70 |
°C |
Storage Temperature |
10~95% RH |
-40 |
|
85 |
°C |
Operating Humidity |
non-condensing |
0 |
|
95% |
RH |
Storage Humidity |
|
0 |
|
95% |
RH |
Electro Static Discharge |
Air Discharge, IEC61000-4-2 |
|
|
15 |
KV |
Electro Static Discharge |
Contact Discharge, IEC61000-4-2 |
|
|
8 |
KV |
Mean Time Between Failure |
Operation Temperature at 25°C, Calculated per MIL-HDBK-217F |
200K |
|
|
h |
Operating Altitude (Elevation) |
All Condition |
|
|
3000 |
m |
Vibration |
10~500Hz,10min./1cycle, 60min.each along X, Y, Z axes |
|
|
5 |
G |
Surge Voltage |
Line-Neutral |
|
|
1 |
KV |
Surge Voltage |
Line-PE & Neutral-PE |
|
|
2 |
KV |
Mechanical Drawing and Technical Chart |
Note:
|