Specifications
3 Year Warranty |
|
Approvals: |
Single Output |
||||
Model Number |
Output Voltage |
Max. Output Current |
Total Regulation |
Maximum Output Power |
SHWA15-S01 |
3 ~ 4.99 VDC |
1.44 ~ 2.40 A |
±5% |
7.2W |
SHWA15-S02 |
5 ~ 5.99 VDC |
2.00 ~ 2.40 A |
±5% |
12W |
SHWA15-S03 |
6.5 ~ 8 VDC |
1.50 ~ 1.84 A |
±5% |
12W |
SHWA15-S04 |
8 ~ 11 VDC |
1.22 ~ 1.68 A |
±5% |
13.5W |
SHWA15-S05 |
11 ~ 13 VDC |
1.15 ~ 1.36 A |
±5% |
15W |
SHWA15-S06 |
13 ~ 16 VDC |
0.94 ~ 1.15 A |
±5% |
15W |
SHWA15-S07 |
16 ~ 21 VDC |
0.72 ~ 0.94 A |
±5% |
15W |
SHWA15-S08 |
21 ~ 27 VDC |
0.55 ~ 0.72 A |
±5% |
15W |
SHWA15-S09 |
27 ~ 33 VDC |
0.45 ~ 0.55 A |
±3% |
15W |
SHWA15-S10 |
33 ~ 40 VDC |
0.37 ~ 0.45 A |
±3% |
15W |
SHWA15-S11 |
40 ~ 48 VDC |
0.32 ~ 0.37 A |
±3% |
15W |
SHWA15-S12 |
48 ~ 55 VDC |
0.28 ~ 0.32 A |
±3% |
15W |
SHWA15-S02~S04 required to use AWG#18/4FT output cable.
SHWA15-S05~S09 required to use AWG#20/4FT output cable.
SHWA15-S10~S11 required to use AWG#22/4FT output cable.
The regulation and efficiency will be changed by modified output cable.
Electrical Characteristics |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Safety Approvals Input Voltage Range |
Safety Approval & Specification in Label |
100 |
|
240 |
VAC |
Operate Voltage Range |
Derate linearly from 100% load at 90VAC to 80% load at 80VAC |
80 |
|
275 |
VAC |
Input Frequency |
Sine wave |
47 |
|
63 |
Hz |
Output Power Range |
See Rating Chart |
|
|
15 |
W |
Low Line Input Current |
Full load, Vin=100VAC |
|
|
0.40 |
A |
High Line Input Current |
Full load, Vin=240VAC |
|
|
0.20 |
A |
Low Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=100VAC |
|
|
50 |
A |
High Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=240VAC |
|
|
115 |
A |
Efficiency |
Full Load, Vin=230VAC |
71 |
|
84 |
% |
Line Regulation |
Full Load, Vin=100~120VAC or 200~240VAC |
|
|
1 |
% |
Over Load Protection |
Recovers automatically after fault condition is removed |
110 |
|
150 |
% |
Time of Transient Response |
Io=Full Load to Half Load, Vin=110VAC |
|
|
4 |
ms |
Hold-Up Time |
Full Load, Vin=110VAC |
|
|
10 |
ms |
Start Up Time |
Full Load, Vin=100~240VAC |
|
|
3 |
s |
* Ripple & Noise (Peak to Peak) |
|
|
1 |
2 |
% |
Temperature Coefficient |
All output |
|
|
±0.04 |
%/°C |
Dielectric Withstanding Voltage(P-S) |
Primary to Secondary, limit current<10mA |
|
|
4000 |
VAC |
EMC Emission |
Compliance to EN55011(CISPR11), EN60601-1-2 |
B |
|
|
Class |
* 3~5V output model max ripple 2%.
Environmental |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Operating Temperature |
Derate linearly from 100% load at 40°C to 50% load at 70°C |
-10 |
|
70 |
°C |
Storage Temperature |
10~95% RH |
-40 |
|
85 |
°C |
Operating Humidity |
non-condensing |
0 |
|
95% |
RH |
Storage Humidity |
|
0 |
|
95% |
RH |
Electro Static Discharge |
Air Discharge, IEC61000-4-2 |
|
|
15 |
KV |
Electro Static Discharge |
Contact Discharge, IEC61000-4-2 |
|
|
8 |
KV |
Mean Time Between Failure |
Operation Temperature at 25°C, Calculated per MIL-HDBK-217F |
200K |
|
|
h |
Operating Altitude (Elevation) |
All Condition |
|
|
3000 |
m |
Vibration |
10~500Hz,10min./1cycle, 60min.each along X, Y, Z axes |
|
|
5 |
G |
Surge Voltage |
Line-Neutral |
|
|
1 |
KV |
Surge Voltage |
Line-PE & Neutral-PE |
|
|
2 |
KV |
Mechanical Drawing and Technical Chart |
Note:
|
Output Derating Chart: 1. Operating Temperature: -10 to 70°C 2. Derate linearly from 100% load at 40°C to 50% load at 70°C |
Datasheet PDF