Print this page

SHWA15 SERIES


15W Multi-Plug Wall Mount for Medical Equipment


  • Wide Operating Voltage 80 to 275VAC, 47 to 63Hz
  • Interchangeable AC plug(USA, UK, Europe, Australia)
  • Single Output
  • Input to Output : 2MOPP
  • High ESD immunity
  • Suitable home healthcare environment
  • Suitable professional healthcare facility

Specifications

 

 

3 Year Warranty

 

Approvals:

 

Single Output

Model Number

Output Voltage

Max. Output Current

Total Regulation

Maximum Output Power

SHWA15-S01

3 ~ 4.99 VDC

1.44 ~ 2.40 A

±5%

7.2W

SHWA15-S02

5 ~ 5.99 VDC

2.00 ~ 2.40 A

±5%

12W

SHWA15-S03

6.5 ~ 8 VDC

1.50 ~ 1.84 A

±5%

12W

SHWA15-S04

8 ~ 11 VDC

1.22 ~ 1.68 A

±5%

13.5W

SHWA15-S05

11 ~ 13 VDC

1.15 ~ 1.36 A

±5%

15W

SHWA15-S06

13 ~ 16 VDC

0.94 ~ 1.15 A

±5%

15W

SHWA15-S07

16 ~ 21 VDC

0.72 ~ 0.94 A

±5%

15W

SHWA15-S08

21 ~ 27 VDC

0.55 ~ 0.72 A

±5%

15W

SHWA15-S09

27 ~ 33 VDC

0.45 ~ 0.55 A

±3%

15W

SHWA15-S10

33 ~ 40 VDC

0.37 ~ 0.45 A

±3%

15W

SHWA15-S11

40 ~ 48 VDC

0.32 ~ 0.37 A

±3%

15W

SHWA15-S12

48 ~ 55 VDC

0.28 ~ 0.32 A

±3%

15W

SHWA15-S02~S04 required to use AWG#18/4FT output cable.

SHWA15-S05~S09 required to use AWG#20/4FT output cable.

SHWA15-S10~S11 required to use AWG#22/4FT output cable.

The regulation and efficiency will be changed by modified output cable.

 

Electrical Characteristics

Parameter

Test Conditions

Min.

Typ.

Max.

Unit

Safety Approvals Input Voltage Range

Safety Approval & Specification in Label

100

 

240

VAC

Operate Voltage Range

Derate linearly from 100% load at 90VAC to 80% load at 80VAC

80

 

275

VAC

Input Frequency

Sine wave

47

 

63

Hz

Output Power Range

See Rating Chart

 

 

15

W

Low Line Input Current

Full load, Vin=100VAC

 

 

0.40

A

High Line Input Current

Full load, Vin=240VAC

 

 

0.20

A

Low Line Input Inrush Current

Full load, 25°C, Cool start, Vin=100VAC

 

 

50

A

High Line Input Inrush Current

Full load, 25°C, Cool start, Vin=240VAC

 

 

115

A

Efficiency

Full Load, Vin=230VAC

71

 

84

%

Line Regulation

Full Load, Vin=100~120VAC or 200~240VAC

 

 

1

%

Over Load Protection

Recovers automatically after fault condition is removed

110

 

150

%

Time of Transient Response

Io=Full Load to Half Load, Vin=110VAC

 

 

4

ms

Hold-Up Time

Full Load, Vin=110VAC

 

 

10

ms

Start Up Time

Full Load, Vin=100~240VAC

 

 

3

s

* Ripple & Noise (Peak to Peak)

 

 

1

2

%

Temperature Coefficient

All output

 

 

±0.04

%/°C

Dielectric Withstanding Voltage(P-S)

Primary to Secondary, limit current<10mA

 

 

4000

VAC

EMC Emission

Compliance to EN55011(CISPR11), EN60601-1-2

B

 

 

Class

* 3~5V output model max ripple 2%.

 

  

Environmental

Parameter

Test Conditions

Min.

Typ.

Max.

Unit

Operating Temperature

Derate linearly from 100% load at 40°C to 50% load at 70°C

-10

 

70

°C

Storage Temperature

  10~95% RH

-40

 

85

°C

Operating Humidity

  non-condensing

0

 

95%

RH

Storage Humidity

 

0

 

95%

RH

Electro Static Discharge

Air Discharge, IEC61000-4-2

 

 

15

KV

Electro Static Discharge

Contact Discharge, IEC61000-4-2

 

 

8

KV

Mean Time Between Failure

Operation Temperature at 25°C, Calculated per MIL-HDBK-217F

200K

 

 

h

Operating Altitude (Elevation)

All Condition

 

 

3000

m

Vibration

10~500Hz,10min./1cycle, 60min.each along X, Y, Z axes

 

 

5

G

Surge Voltage

  Line-Neutral

 

 

1

KV

Surge Voltage

  Line-PE & Neutral-PE

 

 

2

KV

 

Mechanical Drawing and Technical Chart

          

Note:

  1. Dimensions are shown in mm.
  2. Weight: 165g approx.  
  3. Optional output connector.

Output Derating Chart:

1. Operating Temperature: -10 to 70°C

2. Derate linearly from 100% load at 40°C to 50% load at 70°C

 

Datasheet PDF