Specifications
Â
3 Year Warranty |
 |
Approvals:Â Â |
Â
Single Output |
||||
Model Number |
Output Voltage |
Max. Output Current |
Load Regulation |
Maximum Output Power |
SHWA10D -S01Â |
3 ~ 5 VDC |
1.20 ~ 2.00 A |
±5% |
6W |
SHWA10D -S02Â |
5 ~ 6 VDC |
1.33 ~ 1.60 A |
±5% |
8W |
SHWA10D -S03Â |
6 ~ 8 VDC |
1.00 ~ 1.33 A |
±5% |
8W |
SHWA10D -S04 |
8 ~ 11 VDC |
0.90 ~ 1.25 A |
±5% |
10W |
SHWA10D -S05 |
11 ~ 13 VDC |
0.76 ~ 0.90 A |
±5% |
10W |
SHWA10D -S06 |
13 ~ 16 VDC |
0.62 ~ 0.76 A |
±5% |
10W |
SHWA10D -S07 |
16 ~ 21 VDC |
0.47 ~ 0.62 A |
±5% |
10W |
SHWA10D -S08 |
21 ~ 27 VDC |
0.37 ~ 0.47 A |
±5% |
10W |
SHWA10D -S09 |
27 ~ 33 VDC |
0.30 ~ 0.37 A |
±3% |
10W |
SHWA10D -S10 |
33 ~ 40 VDC |
0.25 ~ 0.30 A |
±3% |
10W |
SHWA10D -S11 |
40 ~ 50 VDC |
0.20 ~ 0.25 A |
±3% |
10W |
SHWA10D -S12 |
50 ~ 55 VDC |
0.19 ~ 0.20 A |
±3% |
10W |
Selected output connectors and wire, please refer to Appendix.
SHWA10D-S01~S05 are required to use AWG#20/4FT output cable.
SHWA10D-S06~S12 are required to use AWG#24/4FT output cable.
The regulation and efficiency will be changed by modified output cable.
Â
Electrical Characteristics |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Safety Approvals Input Voltage Range |
Safety Approval & Specification in Label |
100 |
 |
240 |
VAC |
Operate Voltage Range |
Derate linearly from 100% load at 90VAC to 80% load at 80VAC |
80 |
 |
275 |
VAC |
Input Frequency |
Sine wave |
47 |
 |
63 |
Hz |
Output Power Range |
See Rating Chart |
 |
 |
10 |
W |
Low Line Input Current |
Full load, Vin=100VAC |
 |
 |
0.30 |
A |
High Line Input Current |
Full load, Vin=240VAC |
 |
 |
0.20 |
A |
Low Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=100VAC |
 |
 |
40 |
A |
High Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=240VAC |
 |
 |
100 |
A |
Efficiency |
Full Load, Vin=230VAC |
65 |
 |
82 |
% |
Line Regulation |
Full Load, Vin=100~120VAC |
 |
 |
1 |
% |
Over Load Protection |
Recovers automatically after fault condition is removed |
110 |
 |
150 |
% |
Time of Transient Response |
Io=Full Load to Half Load, Vin=110VAC |
 |
 |
4 |
ms |
Hold-Up Time |
Full Load, Vin=100VAC |
 |
 |
12 |
ms |
Start Up Time |
Full Load, Vin=100~240VAC |
 |
 |
2 |
s |
*Â Ripple & Noise (Peak to Peak) |
 |
 |
1 |
2 |
% |
Temperature Coefficient |
All output |
 |
 |
±0.04 |
%/°C |
Dielectric Withstanding Voltage(P-S) |
Primary to Secondary, limit current<10mA |
 |
 |
4000 |
VAC |
EMC Emission |
Compliance to EN55011(CISPR11), EN60601-1-2 |
B |
 |
 |
Class |
* 3~5V output model max ripple 2%.
Â
   Â
Environmental |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Operating Temperature |
Derate linearly from 100% load at 40â to 50% load at 70â |
-10 |
 |
70 |
°C |
Storage Temperature |
 10~95% RH |
-40 |
 |
85 |
°C |
Operating Humidity |
 non-condensing |
0 |
 |
95% |
RH |
Storage Humidity |
 |
0 |
 |
95% |
RH |
Electro Static Discharge |
Air Discharge, IEC61000-4-2 |
 |
 |
15 |
KV |
Electro Static Discharge |
Contact Discharge, IEC61000-4-2 |
 |
 |
8 |
KV |
Mean Time Between Failure |
Operation Temperature at 25âJ, Calculated per MIL-HDBK-217F |
200K |
 |
 |
h |
Operating Altitude (Elevation) |
All Condition |
 |
 |
3000 |
m |
Vibration |
10~500Hz,10min./1cycle, 60min.each along X, Y, Z axes |
 |
 |
5 |
G |
Surge Voltage |
 Line-Neutral |
 |
 |
1 |
KV |
Surge Voltage |
 Line-PE & Neutral-PE |
 |
 |
2 |
KV |
Â
Mechanical Diagram and Technical Chart |
Â
Note:
  Output Derating Chart:   1. Operating Temperature: -10 to 70°C 2. Derate linearly from 100% load at 40°C  |
Â
Â