Specifications
3 Year Warranty |
|
Approvals: |
Single Output |
|||||||||||
Model Number |
Output Voltage |
Output Current |
Total Regulation |
Maximum Output Power |
Typ. Efficiency |
Typ. No Load Consumption |
|||||
min |
max |
min |
max |
(with fan cooling) |
(Free Air) |
(Forced Air) |
|||||
min |
max |
||||||||||
SHFA100-S05F |
11.0 VDC |
13.0 VDC |
7.69 A |
8.33 A |
11.53 A |
13.63 A |
±3% |
100 W |
150 W |
86% |
0.5 W |
SHFA100-S06F |
13.0 VDC |
16.0 VDC |
6.25 A |
7.69 A |
9.37 A |
11.53 A |
±3% |
100 W |
150 W |
86% |
0.5 W |
SHFA100-S07F |
16.0 VDC |
21.0 VDC |
4.77 A |
6.25 A |
7.14 A |
9.37 A |
±3% |
100 W |
150 W |
87% |
0.5 W |
SHFA100-S08F |
21.0 VDC |
27.0 VDC |
3.70 A |
4.77 A |
5.55 A |
7.14 A |
±3% |
100 W |
150 W |
88% |
0.5 W |
SHFA100-S09F |
27.0 VDC |
33.0 VDC |
3.03 A |
3.70 A |
4.54 A |
5.55 A |
±3% |
100 W |
150 W |
88% |
0.5 W |
SHFA100-S10F |
33.0 VDC |
40.0 VDC |
2.50 A |
3.03 A |
3.75 A |
4.54 A |
±3% |
100 W |
150 W |
88% |
0.5 W |
SHFA100-S11F |
40.0 VDC |
50.0 VDC |
2.00 A |
2.50 A |
3.00 A |
3.75 A |
±3% |
100 W |
150 W |
88% |
0.5 W |
Â
Electrical Characteristics |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Safety Approvals Input Voltage Range |
Safety Approval & Specification in Label |
100 |
|
240 |
VAC |
Operate Voltage Range |
|
90 |
|
260 |
VAC |
Input Frequency |
Sine wave |
47 |
|
63 |
Hz |
Power Factor Correction |
|
0.90 |
|
1 |
|
Output Power Range |
See Rating Chart |
|
100 |
150 |
W |
Low Line Input Current |
Full load, Vin=100VAC |
|
1.4 |
|
A |
High Line Input Current |
Full load, Vin=240VAC |
|
0.7 |
|
A |
Low Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=100VAC |
|
|
50 |
A |
High Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=240VAC |
|
|
100 |
A |
Safety Ground Leakage Current |
Vin=240VAC, Fi=60Hz |
|
|
0.25 |
mA |
Efficiency |
Full Load, Vin=230VAC |
86 |
|
88 |
% |
Line Regulation |
Full Load, Vin=100~120VAC or 200~240VAC |
|
|
1 |
% |
Over Voltage Protection |
|
112 |
|
132 |
% |
Over Load Protection |
Recovers automatically after fault condition is removed |
160 |
|
200 |
% |
Time of Transient Response |
Io=Full Load to Half Load, Vin=110VAC |
|
|
4 |
ms |
Hold-Up Time |
Full Load, Vin=100VAC |
|
|
16 |
ms |
Start Up Time |
Full Load, Vin=100~240VAC |
|
|
0.5 |
s |
Insulation Resistance |
|
50 |
|
|
MΩ |
Ripple & Noise (Peak to Peak) |
|
|
|
1 |
% |
Temperature Coefficient |
All output |
|
|
±0.04 |
%/°C |
Dielectric Withstanding Voltage(P-S) |
Primary to Secondary, limit current<10mA |
4000 |
|
|
VAC |
Dielectric Withstanding Voltage(P-G) |
Primary to PE, limit current<10mA |
2828 |
|
|
VAC |
EMC Emission |
Compliance to EN55011(CISPR11), EN60601-1-2 |
B |
|
|
Class |
Environmental |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Operating Temperature |
Derate linearly from 100% load at 40â to 50% load at 70â |
-10 |
|
70 |
°C |
Storage Temperature |
10~95% RH |
-40 |
|
85 |
°C |
Operating Humidity |
non-condensing |
0 |
|
95% |
RH |
Storage Humidity |
|
0 |
|
95% |
RH |
Electro Static Discharge |
Air Discharge, IEC61000-4-2 |
|
|
15 |
KV |
Electro Static Discharge |
Contact Discharge, IEC61000-4-2 |
|
|
8 |
KV |
Mean Time Between Failure |
Operation Temperature at 25âJ, Calculated per MIL-HDBK-217F |
200K |
|
 |
h |
Operating Altitude (Elevation) |
All Condition |
|
|
3000 |
m |
Vibration |
10~500Hz,10min./1cycle, 60min.each along X, Y, Z axes |
|
|
5 |
G |
Surge Voltage |
Line-Neutral |
|
|
1 |
KV |
Surge Voltage |
Line-PE & Neutral-PE |
|
|
2 |
KV |
Mechanical Diagram and Technical Chart |
Note: 1. Dimensions are shown in mm & inch 2. Weight: approx. 180~250g 3. Input connector mates with JST housing 4. Output connector mates with JST housing
PIN Chart
|
Output Derating Curve
|
1. Operating Temperature: -10 to 70°C 2. Derate linearly from 100% load at 40°C to 50% load at 70°C |
Datasheet PDF