Specifications
3 Year Warranty |
|
Approvals: |
Single Output |
|||||||
Product Number |
Output Voltage |
Max. Output Current |
Total Regulation |
Typ.Efficiency (%) |
Ripple & Noise (mVp-p) |
Max. Output Power |
Pno (max.) |
SHFA58-S02 |
5 ~ 6 VDC |
5.50 ~ 6.60 A |
±5% |
80 |
100 |
33W |
0.5W |
SHFA58-S03 |
6 ~ 8 VDC |
5.00 ~ 6.66 A |
±5% |
82 |
100 |
40W |
0.5W |
SHFA58-S04 |
8 ~ 11 VDC |
4.54 ~ 6.25 A |
±5% |
86 |
100 |
50W |
0.5W |
SHFA58-S05 |
11 ~ 13 VDC |
4.61 ~ 5.45 A |
±5% |
87 |
100 |
60W |
0.5W |
SHFA58-S06 |
13 ~ 16 VDC |
3.75 ~ 4.67 A |
±5% |
87 |
100 |
60W |
0.5W |
SHFA58-S07 |
16 ~ 21 VDC |
2.85 ~ 3.75 A |
±5% |
87 |
100 |
60W |
0.5W |
SHFA58-S08 |
21 ~ 27 VDC |
2.22 ~ 2.85 A |
±3% |
88 |
100 |
60W |
0.5W |
SHFA58-S09 |
27 ~ 33 VDC |
1.81 ~ 2.22 A |
±3% |
88 |
200 |
60W |
0.5W |
SHFA58-S10 |
33 ~ 40 VDC |
1.50 ~ 1.81 A |
±3% |
89 |
200 |
60W |
0.5W |
SHFA58-S11 |
40 ~ 50 VDC |
1.20 ~ 1.50 A |
±3% |
90 |
200 |
60W |
0.5W |
SHFA58-S12 |
50 ~ 55 VDC |
1.09 ~ 1.20 A |
±3% |
90 |
200 |
60W |
0.5W |
Electrical Characteristics |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Safety Approvals Input Voltage Range |
Safety Approval & Specification in Label |
100 |
|
240 |
VAC |
Operate Voltage Range |
Derate linearly from 100% load at 90VAC to 80% load at 80VAC |
80 |
|
275 |
VAC |
Input Frequency |
Sine wave |
47 |
|
63 |
Hz |
Output Power Range |
See Rating Chart |
|
|
60 |
W |
Low Line Input Current |
Full load, Vin=100VAC |
|
|
1.2 |
A |
High Line Input Current |
Full load, Vin=240VAC |
|
|
0.6 |
A |
Low Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=100VAC |
|
|
30 |
A |
High Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=240VAC |
|
|
60 |
A |
Safety Ground Leakage Current |
Vin=240VAC, Fi=60Hz |
|
|
0.25 |
mA |
Efficiency |
Full Load, Vin=230VAC |
80 |
|
90 |
% |
Line Regulation |
Full Load, Vin=100~120VAC or 200~240VAC |
|
|
1 |
% |
Over Voltage Protection |
|
110 |
|
275 |
% |
Over Load Protection |
Recovers automatically after fault condition is removed |
110 |
|
150 |
% |
Time of Transient Response |
Io=Full Load to Half Load, Vin=110VAC |
|
|
4 |
ms |
Hold-Up Time |
Full Load, Vin=100VAC |
|
|
12 |
ms |
Start Up Time |
Full Load, Vin=100~240VAC |
0.25 |
|
0.5 |
s |
Insulation Resistance |
|
50 |
|
|
MΩ |
Ripple & Noise (Peak to Peak) |
|
|
|
1 |
% |
Temperature Coefficient |
All output |
|
|
±0.04 |
%/°C |
Dielectric Withstanding Voltage(P-S) |
Primary to Secondary, limit current<10mA |
|
|
4000 |
VAC |
Dielectric Withstanding Voltage(P-G) |
Primary to PE, limit current<10mA |
|
|
1500 |
VAC |
EMC Emission |
Compliance to EN55011(CISPR11), EN60601-1-2 |
B |
|
|
Class |
Environmental |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Operating Temperature |
Derate linearly from 100% load at 40â to 50% load at 70â |
-10 |
|
70 |
°C |
Storage Temperature |
10~95% RH |
-40 |
|
85 |
°C |
Operating Humidity |
non-condensing |
0 |
|
95% |
RH |
Storage Humidity |
|
0 |
|
95% |
RH |
Electro Static Discharge |
Air Discharge, IEC61000-4-2 |
|
|
15 |
KV |
Electro Static Discharge |
Contact Discharge, IEC61000-4-2 |
|
|
8 |
KV |
Mean Time Between Failure |
Operation Temperature at 25âJ, Calculated per MIL-HDBK-217F |
200K |
|
|
h |
Operating Altitude (Elevation) |
All Condition |
|
|
3000 |
m |
Vibration |
10~500Hz,10min./1cycle, 60min.each along X, Y, Z axes |
|
|
5 |
G |
Surge Voltage |
Line-Neutral |
|
|
1 |
KV |
Surge Voltage |
Line-PE & Neutral-PE |
|
|
2 |
KV |
Mechanical Diagram and PIN Chart |
Note:
09-50-3031/35977-0300 and Molex 2478/ 35922 series crimp terminal
housing 09-50-3061/35977-0600 and Molex 2478/ 35922 series crimp terminal.
PIN CHART
|
Output Derating Curve |