Specifications
5 Year Warranty |
|
Approvals: |
Single Output |
||||
Product Number |
Output Voltage |
Max. Output Current |
Total Regulation |
Maximum Output Power |
SMDA130-S05 |
12 VDC |
10.84 A |
5% |
130W |
SMDA130-S08 |
24 VDC |
5.42 A |
3% |
130W |
Total Regulation is guaranteed by below configuration.
SMDA130-S05 is required to use AWG16*5C/4FT output cable.
SMDA130-S08 is required to use AWG18*4C/4FT output cable.
Electrical Characteristics |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Safety Approvals Input Voltage Range |
Safety Approval & Specification in Label |
100 |
|
240 |
VAC |
Operate Voltage Range |
|
90 |
|
260 |
VAC |
Input Frequency |
Sine wave |
47 |
|
63 |
Hz |
Power Factor Correction |
|
0.95 |
|
1 |
|
Output Power Range |
See Rating Chart |
|
|
130 |
W |
Low Line Input Current |
Full load, Vin=100VAC |
|
|
1.58 |
A |
High Line Input Current |
Full load, Vin=240VAC |
|
|
0.64 |
A |
Low Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=100VAC |
|
|
30 |
A |
High Line Input Inrush Current |
Full load, 25°C, Cool start, Vin=240VAC |
|
|
50 |
A |
Safety Ground Leakage Current |
Vin=240VAC, Fi=60Hz |
|
|
0.1 |
mA |
Efficiency |
Full Load, Vin=230VAC |
86.1 |
|
88.7 |
% |
Line Regulation |
Full Load, Vin=100~120VAC or 200~240VAC |
0.5 |
|
1 |
% |
Over Voltage Protection |
|
112 |
|
132 |
% |
Over Load Protection |
Recovers automatically after fault condition is removed |
110 |
|
150 |
% |
Time of Transient Response |
Io=Full Load to Half Load, Vin=110VAC |
|
|
4 |
ms |
Hold-Up Time |
Full Load, Vin=100VAC |
|
|
16 |
ms |
Start Up Time |
Full Load, Vin=100~240VAC |
|
|
2 |
s |
Insulation Resistance |
|
50 |
|
|
MΩ |
Ripple & Noise (Peak to Peak) |
|
|
|
1 |
% |
Temperature Coefficient |
All output |
|
|
±0.04 |
%/°C |
Dielectric Withstanding Voltage(P-S) |
Primary to Secondary, limit current<10mA |
|
|
4000 |
VAC |
Dielectric Withstanding Voltage(P-G) |
Primary to PE, limit current<10mA |
|
|
1500 |
VAC |
EMC Emission |
Compliance to EN55011(CISPR11), EN60601-1-2 |
B |
|
|
Class |
Environmental |
|||||
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
Operating Temperature |
Derate linearly from 100% load at 40℃ to 50% load at 70℃ |
-10 |
|
70 |
°C |
Storage Temperature |
10~95% RH |
-40 |
|
85 |
°C |
Operating Humidity |
non-condensing |
0 |
|
95% |
RH |
Storage Humidity |
|
0 |
|
95% |
RH |
Electro Static Discharge |
Air Discharge, IEC61000-4-2 |
|
|
15 |
KV |
Electro Static Discharge |
Contact Discharge, IEC61000-4-2 |
|
|
8 |
KV |
Mean Time Between Failure |
Operation Temperature at 25℃J, Calculated per MIL-HDBK-217F |
200K |
|
|
h |
Operating Altitude (Elevation) |
All Condition |
|
|
3000 |
m |
Vibration |
10~500Hz,10min./1cycle, 60min.each along X, Y, Z axes |
|
|
5 |
G |
Surge Voltage |
Line-Neutral |
|
|
1 |
KV |
Surge Voltage |
Line-PE & Neutral-PE |
|
|
2 |
KV |
Mechanical Diagram |
Note: 1. Dimensions are shown in mm. 2. Weight: 778-800g approx. 3. Optional output connector. |
Derating Chart |
1. Operating Temperature: -10 to 70¢XC 2. Derate linearly from 100% load at 40¢XC to 50% load at 70¢XC
|